WARNING
Contact bars must be inserted all the way into the meter or you
may get an inaccurate test result.
Every time you perform a control solution test, you must enter
into the “QC” test mode and then the test result with “QC” symbol
will be stored in the meter memory. Failure to do so will confuse
the blood glucose test result with the control solution test result in
memory.
The control solution test with QC symbol can be stored in meter
memory, but not be calculated to the average results.